Function features

1. Microprocessor based servo controlled temperature system

2. Alarms functions: Power failure; Over temperature; Temperature deviation; Temperature sensor failure; Fan failure;

3. Humidity is adjustable continuously

4. A keypad lock prevent inadvertent changes the setting

5. Integral timer to record therapy time

6. Skin temperature sensor failure protect function avoid over temperature

7. Air temperature, skin temperature, timer and heat power separately displayed by LED

8. >37 centigrade temperature setting

9. Silence/reset button for alarms

10. Trough for pulling and pushing, which is convenient to add water and wash

11. Fan with low noise and long life-span

12. Temperature deviation and over temperature could be amended in front panel directly

13. A second thermal cut-off function for more safety

14. The infant bed inclination angle is adjustable

15. Six operating windows, two sides operating windows can be swiveled

16. Heater is electrically cut-out if air temperature inside incubator exceed 40 centigrade

 

Standard configuration

Main body(including the transparent hood, control system, infant bed, water trough), skin temperature sensor, I. V. Pole, mattress, cabinet, castors

 

Optional parts: Phototherapy up side unit; Phototherapy from up and down sides; RS232(485)conncector; Tray/pole;

Infant bed pulled from side door; Infant bed pulled from side and front door; Double wall hood

 

Technical data:

1. Wer supply: AC 110/220V 60/50Hz

2. Wer input: 420VA

 

Operating conditions:

1. Evironment temperature: 20~30certigrade

2. Evironment relative humidity: 30%~75%RH

3. Evironment air velocity of flow: <0.3m/s

4. Air Temperature control range: 25~38certigrade

5. Skin temperature control range: 32~38.5certigrade

6. Skin temperature sensor accuracy: ±0.3certigrade

7. Temperature fluctuation: ±0.5certigrade

8. Uniformity of mattress temperature: 0.8certigrade

9. Warm-up time: 45min

10. Internal noise level: 55dB

11. Infant bed tilt angle: ±10degree

12. Mattress size: 65cm(L)*37cm(W)

13. Trough capacity: 1200ml

14. Air filter: 0.5um

 

Package:

Each unit is packed in two wooden crate

Size (control unit and cabinet):

12*70* 96CM

N. W.: 80KG G. W.: 110KG

Hood: 97*56*63CM

N. W.: 15KG G. W.: 40KG

 

Transport and storage

Environment Temp: -10--+70certigrade

Environment Relative Humidity: 80%

Atmospheric Pressure: 500~1060Hpa